      Series 5000

  Model 5300/Advanced Lithium Cell Formation System

                  

 Performance Specifications

    
  System Size:                              From 256 to > 100,000 cells positions
  Measurement Voltage Range:       0 - 5.0 Volts
  Maximum Charge Voltage:           5.0 Volts
  Minimum Discharge Voltage:        2.0 Volts
  Voltage Measurement Accuracy:   +/- 5 milliVolts
  Voltage Control Accuracy:            +/- 5 milliVolts
  Ripple:                                       0 (no measurable ripple)
  Maximum Charge Current:            2.5 Amps (limited to 400 Amps total for each 256

                                                   cell cabinet)
  Maximum Discharge Current:        2.5 Amps
  Current Setting Range:                 0.01 to 2.5 Amps (In CV mode, the operating Current                                                    can go to 0 milliAmps)
  Current Measurement Accuracy:    +/- 2.5 milliAmps
  Current Control Accuracy:             +/- 2.5 milliAmps
  Capacity Accuracy:                      +/- 2.5 milliAmps-hours at the C rate
  Time:                                          Update Time 500 milliSeconds
  Data Record Interval                     1 minute or multiple thereof
  Programming Time                        hh:mm
  Data File Display Time                   hh:mm:ss
  Digital I/O:                                   7 inputs and 8 outputs available for control purposes,                                                      for each set of 256 cells  
  Cell Type:                                    Lithium Ion or Lithium Polymer
  AC Power Supply:                        180 - 264 V or 360 - 500 V, 3 phase, 47- 63 Hz.2.5 KW                                                     per 256 cell Cabinet
  UPS:                                           Required for PC computer(s) only, typically  1400 VA.
  Cooling Air:                                 12 cubic meters per minute per 256 cell cabinet
  Instrument Air:                             3 to 4 Bar with minimal volume
  Environmental Temperature:          5C to 35C
  Environmental Temperature:          RH 20% to 80% non-condensing

 